Technical Data
Input Voltage: Uniphase AC 100~240V,±10%
Supply Frequency: 50/60Hz,±1Hz
Input Power: 286W
Output Power: 25.2W
Charge voltage: DC 25.2V
Charge current: 1.0A
Nominal value of the focal point: 0.8mm
Anode angle: 19º
Available maximum tube voltage at 2mA: 60kV
Maximum output electric power: 0.12kW
Maximum loading factor combination: 60kV、2mA、2s
Repeatability of radiation output: No more than 0.5
Tube voltage (The deviation is within the range of ±10%): 60kV
Tube current (The deviation is within the range of ±20%): 2mA
Running mode: Discontinuous operation, maximum excitation (open) time : minimum excitation (close) time = 1 : 30
Total filtration of X-ray tube assembly: >2.5mm Al
Intrinsic filtration of X-ray tube assembly: 0.75mm Al
Additional filtration: 1.00mm Al
Half-value layer: ≥1.5mm Al(60kV)
Leakage radiation: ≤0.25mGy/h(60kV, 2mA, 2s)
Adjustment range of the loading time: 0.01~2.00s, adjustable
Distance of the Focal point to skin (The deviation is within 0~5%): 211mm±3mm
X-ray field size at the end of beam limiting tube: round,Φ 56mm±3mm